LOW-LEVEL DISTORTION ANALYSIS TECHNIQUES:Every measurement system has some finite limit on the range of values it can handle.Consider the needs of a system to measure the distortion of an audio power amplifier. Typical modern amplifiers may have distortion levels (harmonic or intermodulation) that are below the noise floor of your ADC board. Even with waveform averaging to reduce the noise in the measurement, you may find that distortion from the board itself is larger than anything from the amplifier. In other words, if you send a high-level stimulus signal to the amplifier (or other low-distortion system) under test, then at its output will appear that same signal (perhaps greatly amplified), plus one or more much-smaller distortion components. If you want to analyze this composite signal with an ADC board, you must provide attenuation such that the large stimulus portion will not overload the board's input. Even at levels below outright overload, the stimulus portion may induce detectable distortion from the board. If you reduce the level in order to reduce this added distortion, you will also be reducing the amplifier distortion components you are trying to measure. What is needed is a way to reduce the high-level stimulus portion, while leaving the low-level portion unaffected. One standard method is to apply a "notch" filter to the signal, carefully adjusted to cut only the stimulus frequency. For measuring intermodulation distortion, you will need a notch filter for each of the stimulus frequencies. Changing to new stimulus frequencies requires re-tuning. Since the notch must be very sharp in order to cut only the stimulus without affecting nearby difference tones, small tuning errors can allow large amounts of stimulus to pass. A preferred alternate method is to use simple subtraction to remove the stimulus component(s) from the amplifier output signal. This can be done because you have access to both the input and output signals for the amplifier. If you attenuate the output to compensate for the amplifier gain, it should be identical to the input except for any added distortion. You can then subtract them via a simple difference circuit, also known as a "differential amplifier", and obtain the distortion components alone. .-------. To Source >--.--| AMP |--.----------------------> Amplifier | `-------' | Load | .---^--. | | ATTN | | `---.--' | | .-------. .-------. | `--| PHASE |--| + | | `-------' | DIFF |----> To ADC `---------------------------| - | Board `-------'If it happens that the amplifier inverts the signal anyway, you could replace the difference circuit with a passive resistive summing network, like the one previously shown for adding two sources to measure intermodulation distortion. The attenuator can be a simple "volume control" added to the input of the difference circuit. For finer adjustment of the control, use series resistors Ra and Rg to supply the bulk of the attenuation when you know the amplifier gain A. From Amp >------Ra-------. | | .---------. | | |---' To | Control |-----------> Diff | Rc |---. `---------' | Rg | Gnd >---------------^------->A multiple-turn control will give even finer adjustment. If the control has a total resistance of Rc, select Rg about 10 times as big, then select Ra = Rg × A. For example, suppose the amplifier has a gain A = 20 and the control has Rc = 1K. Then select Rg = 10 × Rc = 10 × 1K = 10K Rs = Rg × A = 10K × 20 = 200KThis will allow adjustment for a range of gains from about 19 to 21. Observe the spectrum of the output from the difference circuit while you adjust the attenuator for a minimum at the stimulus frequency. If there are two stimulus frequencies, both peaks should go down at the same time. Note that the difference stage may be driven into clipping at first, before you find the right adjustment, which will create lots of spurious spectral peaks. Once the stimulus itself is cancelled, the subtractor output can be preamplified to drive the ADC for better sensitivity. Consider that if you use the subtractor to null out 48 dB of stimulus (typical, though you can often do even better), you can effectively add that 48 dB to the measurement range of the ADC. That's equivalent to adding another 8 bits to the resolution of your ADC, not counting what you might gain by waveform averaging!
AVERAGING NOTE: For averaging multiple tones produced by STIM3A, the Trigger Source should be set to Stim, which triggers on the start of the stimulus. Internal mode looks only at slope and level, so it would not synchronize to the complex waveform. Some amplifiers, especially at high frequencies, may introduce phase shifts that prevent this simple subtraction method from reducing the stimulus tones to near zero. If they are still big enough to cause ADC intermodulation, you will need to add compensating phase shifts to your subtraction circuit. You can use simple high-pass or low-pass R-C filters with adjustable R (since these produce 90 degree shifts near their cuttoffs), as long as you recognize that these are also changing the effective attenuation. Adjustment can be tricky until you get near the desired null point, since you will be adjusting both phase and level controls together. The phase adjustment could be done on either subtractor input by changing from high-pass to low-pass or vice-versa. Since you typically won't know ahead of time just how much phase shift you need, or in what direction, you must be prepared to switch things around or change capacitor values. If the phase shift is different for each of the two stimulus tones in an intermodulation test, and you can't get satisfactory results by "splitting the difference" on a single phase adjustment, you will need to use a separate source for each stimulus tone. Combine them directly (using the resistive network previously shown) to drive the amplifier, but provide each with separate phase and level controls before combining to go to the subtractor. These same techniques will work in many other cases where you need to measure extremely small distortions, such that any board would be challenged. The subtraction method will work for any type of distortion measurement, as long as you have access to both the input and output signals of the system under test. You can measure extremely low levels of distortion with this approach, much lower than the residual distortion present in the stimulus signal itself. That's because the stimulus distortion is also subtracted away, so all you are left with is whatever was added by the device under test. What's more, this method is not limited to simple tone stimuli, or even steady-state conditions. As long as you can adequately control for phase shifts, the stimulus can be a complex or transient signal, even speech or music.
CREATING LOW IMD ACOUSTIC SIGNALS:To measure distortion in acoustic systems, you need a good microphone with known low distortion of its own. But to measure the harmonic distortion of a microphone, you would need a pure acoustic wave with no distortion. Yet how would you know the acoustic wave's distortion, unless you already had a known microphone?The solution is to use intermodulation measurements instead of harmonic distortion measurements, since harmonic distortion of the individual sources does not compromise the results. The best way to get two tones with no intermodulation is to create them separately via two different loudspeakers or other drivers and let them combine acoustically. Since air is much more linear than any driver, there will be very little intermodulation. Some small amount may arise, particularly if the sound must be mixed in a closed system, due to sound pressure from one source moving the other. Note that this technique does nothing to reduce harmonic distortion, which is generated separately by each source. The dual-source method is also useful when studying the difference tone distortion products generated by the ear itself. These cause tiny "oto-acoustic emissions" of sound from the inner ear, which can be measured by a sensitive microphone in the ear canal for clinical diagnosis of certain conditions. The procedure is often referred to as "DPOAE" for Distortion Product Oto-Acoustic Emissions.
|
||
GO:
Questions? Comments? Contact us!We respond to ALL inquiries, typically within 24 hrs.INTERSTELLAR RESEARCH: 25 Years of Innovative Instrumentation © Copyright 1999 - 2006 by Interstellar Research |